Drug recall

Achieva 1.5T Conversion, Model Number 781283. Nuclear Magnetic Resonance Imaging System

FDA recall event 82038 · Class II · Philips Medical Systems Nederlands

Key facts

Event ID 82038
Recall numbers Z-0869-2019, Z-0870-2019, Z-0871-2019, Z-0872-2019, Z-0873-2019, Z-0874-2019, Z-0875-2019, Z-0876-2019, Z-0877-2019, Z-0878-2019, Z-0879-2019, Z-0880-2019 and 52 more
Recalling firm Philips Medical Systems Nederlands
Product type Devices
Products Achieva 1.5T Conversion, Model Number 781283. Nuclear Magnetic Resonance Imaging System, Achieva 1.5T Conversion, Model Number 781346. Nuclear Magnetic Resonance Imaging System, Achieva 1.5T High Performance, Model Number 781196. Nuclear Magnetic Resonance Imaging System, Achieva 1.5T, Model Number 781178. Nuclear Magnetic Resonance Imaging System, Achieva 1.5T, Model Number 781296. Nuclear Magnetic Resonance Imaging System and 59 more
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
Initiation date Dec 21, 2018
Report date Mar 6, 2019
Termination date Nov 19, 2020
City Best
Country Netherlands
FEI number 3002807880

At a glance

  • 10–99 Products in this recall
  • 1–9 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls