Drug recall

iTotal Hip Replacement System, Total Hip Replacement System, Catalog Number HAA-050-D001-020101

FDA recall event 81723 · Class II · Conformis, Inc.

Key facts

Event ID 81723
Recall numbers Z-0731-2019
Recalling firm Conformis, Inc.
Product type Devices
Products iTotal Hip Replacement System, Total Hip Replacement System, Catalog Number HAA-050-D001-020101
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall The stage 1 and 2 reamer instruments used to prepare fixation for the cup in the acetabulum had a -1mm interference fit with acetabulum bone. The expected interference fit is -2mm.
Initiation date Oct 16, 2018
Report date Jan 23, 2019
Termination date Jun 29, 2020
City Billerica
State MA
Country United States
FEI number 3009844603

At a glance

  • 1–9 Products in this recall
  • 1–9 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date: