Drug recall
Luminex Aries M1 System Model: Aries M6V1 Aries Software v1.0 For In Vitro Diagnostic Use in clinical…
FDA recall event 81420 · Class II · Luminex Corporation
Key facts
| Event ID | 81420 |
|---|---|
| Recall numbers | Z-2292-2019 |
| Recalling firm | Luminex Corporation |
| Product type | Devices |
| Products | Luminex Aries M1 System Model: Aries M6V1 Aries Software v1.0 For In Vitro Diagnostic Use in clinical laboratories. Scanner Label/Labeling: JADAK JDK-2330 REV B 170512-017 |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Firm Initiated |
| Reason for recall | An incorrect control board and firmware installed in some of the Hand Held Barcode Scanners could result in sample/patient mismatch or duplicate results (from other patients specimens) for a patient… |
| Initiation date | Jun 29, 2018 |
| Report date | Aug 28, 2019 |
| Termination date | Apr 28, 2021 |
| City | Austin |
| State | TX |
| Country | United States |
| FEI number | 3002524000 |