Drug recall

Sedecal NOVA FA DR System. for diagnostic radiography.

FDA recall event 81377 · Class II · Sedecal S.A.

Key facts

Event ID 81377
Recall numbers Z-0488-2019
Recalling firm Sedecal S.A.
Product type Devices
Products Sedecal NOVA FA DR System. for diagnostic radiography.
Classification Class II
Status Ongoing
Voluntary or mandated Firm Initiated
Reason for recall A safety problem with installation and maintenance could lead to the breakage of the two steel cables that support the equipment to its roof anchor. Due to this, the equipment may fall and cause harm…
Initiation date Jul 18, 2018
Report date Nov 28, 2018
City Algete
Country Spain
FEI number 3002495042

At a glance

  • 1–9 Products in this recall
  • 1–9 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls