Drug recall

AIA-2000 Analyzer, Product code 022100, 022100R, 022101, 022101R The AIA-2000 analyzer and AIA-900 analyzer…

FDA recall event 81190 · Class II · Tosoh Bioscience Inc

Key facts

Event ID 81190
Recall numbers Z-0599-2019, Z-0600-2019
Recalling firm Tosoh Bioscience Inc
Product type Devices
Products AIA-2000 Analyzer, Product code 022100, 022100R, 022101, 022101R The AIA-2000 analyzer and AIA-900 analyzer are capable of performing three methods of immunoassay: an immunoenzymetric (IEMA) or…, AIA-900Analyzer, Product code 022930, 022930R The AIA-2000 analyzer and AIA-900 analyzer are capable of performing three methods of immunoassay: an immunoenzymetric (IEMA) or sandwich immunoassay, a…
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall Tosoh Bioscience has become aware of a potential issue with the AIA-900 and AIA-2000 analyzers that could result in slippage of the incubator unit contained within the system. If the tension of the…
Initiation date Sep 28, 2018
Report date Dec 19, 2018
Termination date Sep 3, 2020
City Grove City
State OH
Country United States
FEI number 3005529799

At a glance

  • 1–9 Products in this recall
  • 1–9 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls