Drug recall
BD Max Instrument, Catalog Number 441916 441927 (refurbished instrument) - Product Usage: The BD MAX System…
FDA recall event 81145 · Class II · Becton Dickinson & Co.
Key facts
| Event ID | 81145 |
|---|---|
| Recall numbers | Z-0312-2019, Z-0313-2019 |
| Recalling firm | Becton Dickinson & Co. |
| Product type | Devices |
| Products | BD Max Instrument, Catalog Number 441916 441927 (refurbished instrument) - Product Usage: The BD MAX System, used in conjunction with appropriate extraction kits and reagents, is capable of automated…, BD Max Instrument Reader Spare Part, Catalog Number 8090369/443364, 44336409 (Repaired Reader spare part) - Product Usage: The BD MAX System, used in conjunction with appropriate extraction kits and… |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Firm Initiated |
| Reason for recall | A trend for false positive Candida glabrata results has been confirmed. The discrepant results impact the Candida glabrata target only, when in the presence of a Trichomonas vaginalis true positive. |
| Initiation date | Sep 24, 2018 |
| Report date | Nov 7, 2018 |
| Termination date | Apr 10, 2020 |
| City | Sparks |
| State | MD |
| Country | United States |
| FEI number | 1119779 |
At a glance
- 1–9 Products in this recall
- 10–99 FDA inspections of this firm (FEI)
Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.
About FDA drug and product recalls
FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.