Drug recall

Terumo 100/120V AC, Advanced Perfusion System Platform (APS) for use in cardiopulmonary bypass procedures…

FDA recall event 80957 · Class II · Terumo Cardiovascular Systems Corporation

Key facts

Event ID 80957
Recall numbers Z-3242-2018, Z-3243-2018
Recalling firm Terumo Cardiovascular Systems Corporation
Product type Devices
Products Terumo 100/120V AC, Advanced Perfusion System Platform (APS) for use in cardiopulmonary bypass procedures. Catalog Number 801763, Terumo 220/240V AC, Advanced Perfusion System Platform (APS) for use in cardiopulmonary bypass procedures. Catalog Number 801764
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall Screws on a Terumo¿ Advanced Perfusion System 1 Power Manager Board (the base unit circuit board) were not torqued to specification. It is possible that a screw which has not been properly torqued can…, Screws on a Terumo¿ Advanced Perfusion System 1 Power Manager Board (the base unit circuit board) were not torqued to specification. It is possible that a screw which has not been properly torqued…
Initiation date Aug 28, 2018
Report date Oct 3, 2018
Termination date Apr 23, 2020
City Ann Arbor
State MI
Country United States
FEI number 1828100

At a glance

  • 1–9 Products in this recall
  • 10–99 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date: