Drug recall

Connecting Bolt, Model 14-442093

FDA recall event 80543 · Class II · Zimmer Biomet, Inc.

Key facts

Event ID 80543
Recall numbers Z-2875-2018
Recalling firm Zimmer Biomet, Inc.
Product type Devices
Products Connecting Bolt, Model 14-442093
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall Product was manufactured from the incorrect material, which shows a potential decrease in the fatigue strength of the connecting bolt that can lead to fracture.
Initiation date Jun 18, 2018
Report date Aug 29, 2018
Termination date Mar 26, 2020
City Warsaw
State IN
Country United States
FEI number 1825034

At a glance

  • 1–9 Products in this recall
  • None FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date: