Drug recall

Hamilton-C1 Ventilator Unit Product Usage: The HAMILTON-C1 ventilator is intended to provide positive pressure…

FDA recall event 80467 · Class II · Hamilton Medical AG

Key facts

Event ID 80467
Recall numbers Z-2795-2018, Z-2796-2018
Recalling firm Hamilton Medical AG
Product type Devices
Products Hamilton-C1 Ventilator Unit Product Usage: The HAMILTON-C1 ventilator is intended to provide positive pressure ventilation support to adults and pediatrics. The device is used in the in the intensive…, Hamilton-C3 Ventilator Unit Product Usage: The HAMILTON-C3 ventilator is intended to provide positive pressure ventilator support to adults, pediatrics, infants and neonates depending on its…
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall An out of specification component on the control board may cause the ventilator to open the inspiratory channel and the expiratory valve. The ventilator will shut down and an emergency buzzer…
Initiation date Mar 9, 2018
Report date Aug 22, 2018
Termination date Mar 18, 2019
City Bonaduz
Country Switzerland
FEI number 3001421318

At a glance

  • 1–9 Products in this recall
  • 1–9 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls