Drug recall
Advance¿ 35LP Low-Profile PTA Balloon Dilatation Catheter; Catalog numbers: PTA5-35-135-6-2.0 and…
FDA recall event 79861 · Class II · Cook Inc.
Key facts
| Event ID | 79861 |
|---|---|
| Recall numbers | Z-1521-2018 |
| Recalling firm | Cook Inc. |
| Product type | Devices |
| Products | Advance¿ 35LP Low-Profile PTA Balloon Dilatation Catheter; Catalog numbers: PTA5-35-135-6-2 .0 and PTA5-35-135-4-4 .0; UDI: 10827002522647 and 10827002522524 Product Usage: The Advance¿ 35LP Low… |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Firm Initiated |
| Reason for recall | Incorrect product labeling. Product labeled as 6mm x 2cm balloon are packaged with a 4mm x 4 cm balloon label and vice-versa. |
| Initiation date | Apr 16, 2018 |
| Report date | May 2, 2018 |
| Termination date | Jun 27, 2019 |
| City | Bloomington |
| State | IN |
| Country | United States |
| FEI number | 1820334 |