MHI-TM2000 Linear Accelerator System Product Usage: MHI-TM2000 Linear Accelerator System is intended for…

FDA recall event 79439 · Class II · Hitachi Ltd., Medical System Operations Group

Key facts

Event ID 79439
Recall numbers Z-1585-2018
Recalling firm Hitachi Ltd., Medical System Operations Group
Product type Devices
Products MHI-TM2000 Linear Accelerator System Product Usage: MHI-TM2000 Linear Accelerator System is intended for radiation therapy of lesions, tumors and conditions anywhere in the body where radiation…
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall Due to a system controller software anomaly, the patient positioning deviation correction may not be applied and may result in the wrong part of the patient being irradiated.
Initiation date Dec 8, 2017
Report date May 9, 2018
Termination date Apr 11, 2019
City Kashiwa
Country Japan
FEI number 3008238760

At a glance

  • 1–9 Products in this recall
  • 1–9 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls