Drug recall

Allura Xper systems with Certeray X-ray Generator located in regions with high mains (480V) voltage. Product…

FDA recall event 79099 · Class II · Philips Electronics North America Corporation

Key facts

Event ID 79099
Recall numbers Z-0716-2018
Recalling firm Philips Electronics North America Corporation
Product type Devices
Products Allura Xper systems with Certeray X-ray Generator located in regions with high mains (480V) voltage. Product Usage: The Allura Xper series is intended for use on human patients to perform: Vascular…
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall There is an increase in the failure rate of certain Anode Drive Units (ADU5) used in these products. Early life failure of the ADUs only occurs when the hospital mains provide 480V to the system. This…
Initiation date Jan 4, 2018
Report date Mar 7, 2018
Termination date Oct 9, 2019
City Andover
State MA
Country United States
FEI number 1218950

At a glance

  • 1–9 Products in this recall
  • None FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls