Drug recall

AutoMate 2550 "High Speed" Catalog Number ODL25255; AutoMate 1250 "Standard" Catalog Number ODL25125…

FDA recall event 78999 · Class II · Beckman Coulter Inc.

Key facts

Event ID 78999
Recall numbers Z-1151-2018
Recalling firm Beckman Coulter Inc.
Product type Devices
Products AutoMate 2550 "High Speed" Catalog Number ODL25255; AutoMate 1250 "Standard" Catalog Number ODL25125. Chemistry analyzer sample processing and sorting system.
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall A misaligned pipette tip sensor poses the remote risk of cross-contamination when failing to cause a no-tip error, resulting in pipetting patient samples without pipette tips.
Initiation date Jan 18, 2018
Report date Mar 28, 2018
Termination date Sep 27, 2023
City Brea
State CA
Country United States
FEI number 2050012

At a glance

  • 1–9 Products in this recall
  • 10–99 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date: