AutoMate 2550 "High Speed" Catalog Number ODL25255; AutoMate 1250 "Standard" Catalog Number ODL25125…
FDA recall event 78999 · Class II · Beckman Coulter Inc.
Key facts
| Event ID | 78999 |
|---|---|
| Recall numbers | Z-1151-2018 |
| Recalling firm | Beckman Coulter Inc. |
| Product type | Devices |
| Products | AutoMate 2550 "High Speed" Catalog Number ODL25255; AutoMate 1250 "Standard" Catalog Number ODL25125. Chemistry analyzer sample processing and sorting system. |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Firm Initiated |
| Reason for recall | A misaligned pipette tip sensor poses the remote risk of cross-contamination when failing to cause a no-tip error, resulting in pipetting patient samples without pipette tips. |
| Initiation date | Jan 18, 2018 |
| Report date | Mar 28, 2018 |
| Termination date | Sep 27, 2023 |
| City | Brea |
| State | CA |
| Country | United States |
| FEI number | 2050012 |
At a glance
- 1–9 Products in this recall
- 10–99 FDA inspections of this firm (FEI)
Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.
About FDA drug and product recalls
FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.