Tosoh¿Automated Enzyme Immunoassay Analyzer AIA-900, Product Catalog No. 022930. Product Usage: The AIA-900…

FDA recall event 78776 · Class II · Tosoh Bioscience Inc

Key facts

Event ID 78776
Recall numbers Z-0491-2018, Z-0492-2018
Recalling firm Tosoh Bioscience Inc
Product type Devices
Products Tosoh¿Automated Enzyme Immunoassay Analyzer AIA-900, Product Catalog No. 022930. Product Usage: The AIA-900 Automated Enzyme Immunoassay System is an automated analyzer for in vitro diagnostic use…, Tosoh¿Sorter 9tray, Product Code 022931and Tosoh¿Sorter 19tray, Product Code 022932. Product Usage: The AIA-900 9tray Sorter and AIA-900 19tray Sorter supports the automation of the AIA-900 systems.
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall This recall has been initiated due to a potential malfunction of the photo interrupter (Pl). The potential malfunction has been attributed to aging and premature deterioration of the Pl collector…
Initiation date Jun 11, 2017
Report date Feb 7, 2018
Termination date Dec 31, 2019
City Grove City
State OH
Country United States
FEI number 3005529799