Drug recall

The ReNew Reusable Grasper and Dissector Tips are to be used with the ReNew Laparoscopic Handpieces and they…

FDA recall event 78754 · Class II · Microline Surgical, Inc.

Key facts

Event ID 78754
Recall numbers Z-0507-2018
Recalling firm Microline Surgical, Inc.
Product type Devices
Products The ReNew Reusable Grasper and Dissector Tips are to be used with the ReNew Laparoscopic Handpieces and they are indicated for cutting, grasping, dissecting, and coagulation of tissue in endoscopic…
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall The heat-shrink insulation tube may split during autoclave sterilization/re-sterilization. If a split occurs, the heat-shrink could fall off during a surgical procedure and if cautery is needed, there…
Initiation date Jul 13, 2017
Report date Feb 14, 2018
Termination date Feb 6, 2020
City Beverly
State MA
Country United States
FEI number 1000137656

At a glance

  • 1–9 Products in this recall
  • 1–9 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date: