Drug recall

LIFEPAK(R) defibrillation electrodes, Catalog # 11101-000016, MIN #3202380-006 The defibrillation electrodes…

FDA recall event 78303 · Class I · Physio-Control, Inc.

Key facts

Event ID 78303
Recall numbers Z-0186-2018, Z-0187-2018
Recalling firm Physio-Control, Inc.
Product type Devices
Products LIFEPAK(R) defibrillation electrodes, Catalog # 11101-000016, MIN #3202380-006 The defibrillation electrodes are used only with LIFEPAK EXPRESS(R) AED, LIFEPAK CR(R) Plus AED, LIFEPAK 1000…, LIFEPAK(R) Electrode Starter Kits, Catalog # 11101-000017, MIN #3202784-009 The defibrillation electrodes are used only with LIFEPAK EXPRESS(R) AED, LIFEPAK CR(R) Plus AED, LIFEPAK 1000 defibrillator…
Classification Class I
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall The infant artwork on the product instructs incorrect pad placement. Also, the artwork is missing the Type BF Patient connection and High Voltage Warning symbols.
Initiation date Oct 27, 2017
Report date Dec 20, 2017
Termination date Jun 12, 2019
City Redmond
State WA
Country United States
FEI number 3015876