Drug recall

IVEA Product Part Number 16100 Model 500A

FDA recall event 78019 · Class II · Firefly Medical, Inc.

Key facts

Event ID 78019
Recall numbers Z-3239-2017
Recalling firm Firefly Medical, Inc.
Product type Devices
Products IVEA Product Part Number 16100 Model 500A
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall The dowel pins (P/N 16082) that fit in the left and right leg pivot castings (P/Ns 16049/16056) can rotate and gradually maneuver out of the leg casting after several hundred deployment and folding…
Initiation date Aug 24, 2017
Report date Oct 11, 2017
Termination date Nov 1, 2019
City Fort Collins
State CO
Country United States
FEI number 3011390406

At a glance

  • 1–9 Products in this recall
  • None FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date: