Medfusion, Syringe Infusion Pump, Software Version 6 (V6), Rx Only, Manufactured by Smiths Medical ASD, Inc…
FDA recall event 77761 · Class II · Smiths Medical ASD Inc.
Key facts
| Event ID | 77761 |
|---|---|
| Recall numbers | Z-0313-2018, Z-0314-2018 |
| Recalling firm | Smiths Medical ASD Inc. |
| Product type | Devices |
| Products | Medfusion, Syringe Infusion Pump, Software Version 6 (V6), Rx Only, Manufactured by Smiths Medical ASD, Inc., 1500 Eureka Park, Lower Pemberton, Ashford, Kent, TN25 4BF, UK. In the administration of…, Medfusion, Syringe Infusion Pump, Software Version V1.1, Rx Only, Manufactured by Smiths Medical ASD, Inc., 1500 Eureka Park, Lower Pemberton, Ashford, Kent, TN25 4BF, UK. In the administration of… |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Firm Initiated |
| Reason for recall | Potential for the syringe barrel clamp mechanism to loosen, decreasing the pump's ability to accurately detect the diameter of the syringe barrel, resulting in the unit to initiate the "Invalid… |
| Initiation date | Aug 22, 2017 |
| Report date | Jan 10, 2018 |
| Termination date | Apr 14, 2025 |
| City | Minneapolis |
| State | MN |
| Country | United States |
| FEI number | 3012307300 |
At a glance
- 1–9 Products in this recall
- 1–9 FDA inspections of this firm (FEI)
Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.
About FDA drug and product recalls
FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.
Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date: