Drug recall

PROMETRA II Programmable Infusion Pump, Implantable Infusion Pump, Catalog Number/REF 13827, Part Number…

FDA recall event 77442 · Class II · Flowonix Medical, Inc.

Key facts

Event ID 77442
Recall numbers Z-2740-2017
Recalling firm Flowonix Medical, Inc.
Product type Devices
Products PROMETRA II Programmable Infusion Pump, Implantable Infusion Pump, Catalog Number/REF 13827, Part Number PL-31790-02. Manufactured by: Flowonix Medical Inc., --- The correction involves the labeling…
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall Flowonix Medical received a report of a patient implanted with the Prometra II Programmable Pump who may have received a fatal drug overdose during an MRI procedure. The Prometra II pump has an…
Initiation date May 22, 2017
Report date Jul 26, 2017
Termination date Jun 28, 2018
City Mount Olive
State NJ
Country United States
FEI number 3006803715

At a glance

  • 1–9 Products in this recall
  • 1–9 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls