Drug recall

Beckman Coulter Synchron AQUA CAL Level 1, Part No. 471288 Product Usage: Used for the system calibration of…

FDA recall event 77364 · Class II · Beckman Coulter Inc

Key facts

Event ID 77364
Recall numbers Z-2730-2017
Recalling firm Beckman Coulter Inc
Product type Devices
Products Beckman Coulter Synchron AQUA CAL Level 1, Part No. 471288 Product Usage: Used for the system calibration of Na, K, Cl, Urea N, Urea, Glucose Creatine, Ca, CO2, and P in UniCel DxC Systems.
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall The firm identified a negative shift of 5mmol/L for sodium recovery in Synchron Control Level 3 when calibrated with AQUA Cal 1.
Initiation date May 15, 2017
Report date Jul 19, 2017
Termination date Sep 9, 2020
City Brea
State CA
Country United States
FEI number 2050011