Drug recall
UniCel DxH 600 Coulter Cellular Analysis System, Catalog No. B23858.
FDA recall event 77145 · Class II · Beckman Coulter Inc.
Key facts
| Event ID | 77145 |
|---|---|
| Recall numbers | Z-2157-2017, Z-2158-2017 |
| Recalling firm | Beckman Coulter Inc. |
| Product type | Devices |
| Products | UniCel DxH 600 Coulter Cellular Analysis System, Catalog No. B23858., UniCel DxH 800 Coulter Cellular Analysis System, Catalog No. 629029, B24465, B24802, B66445, B68304, B63322. |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Firm Initiated |
| Reason for recall | There is a possibility of a data acquisition disruption that may cause an unusual light scatter event pattern for the white blood cell differential that may not have a system flag and/or message. The… |
| Initiation date | Apr 24, 2017 |
| Report date | May 31, 2017 |
| Termination date | May 22, 2018 |
| City | Brea |
| State | CA |
| Country | United States |
| FEI number | 2050012 |