Drug recall

UniCel DxH 600 Coulter Cellular Analysis System, Catalog No. B23858.

FDA recall event 77145 · Class II · Beckman Coulter Inc.

Key facts

Event ID 77145
Recall numbers Z-2157-2017, Z-2158-2017
Recalling firm Beckman Coulter Inc.
Product type Devices
Products UniCel DxH 600 Coulter Cellular Analysis System, Catalog No. B23858., UniCel DxH 800 Coulter Cellular Analysis System, Catalog No. 629029, B24465, B24802, B66445, B68304, B63322.
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall There is a possibility of a data acquisition disruption that may cause an unusual light scatter event pattern for the white blood cell differential that may not have a system flag and/or message. The…
Initiation date Apr 24, 2017
Report date May 31, 2017
Termination date May 22, 2018
City Brea
State CA
Country United States
FEI number 2050012

At a glance

  • 1–9 Products in this recall
  • 10–99 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date: