Drug recall

ADVIA Centaur Systems Insulin Calibrator; Test Code: Calibrator IRI; Catalog Number: 04618899

FDA recall event 76992 · Class II · Siemens Healthcare Diagnostics, Inc

Key facts

Event ID 76992
Recall numbers Z-2085-2017, Z-2086-2017
Recalling firm Siemens Healthcare Diagnostics, Inc
Product type Devices
Products ADVIA Centaur Systems Insulin Calibrator; Test Code: Calibrator IRI; Catalog Number: 04618899, ADVIA Centaur Systems Insulin ReadyPack; Test Code: IRI; Catalog Number: 02230141, for in vitro diagnostic use in the quantitative determination of insulin in serum.
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall Current lots of ADVIA Centaur/XP/XPT/CP Insulin (IRI) assay recover approximately 40% higher than the World Health Organization (WHO) 1st IRP 66/304 standardization based on slope values. Other…
Initiation date Apr 5, 2017
Report date May 24, 2017
Termination date Apr 2, 2019
City East Walpole
State MA
Country United States
FEI number 1219913

At a glance

  • 1–9 Products in this recall
  • 10–99 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls