Drug recall

ADVIA Chemistry Enzyme 2 Calibrator, multi-analyte mixture Product Usage: Assays included in Enzyme 2…

FDA recall event 76695 · Class III · Siemens Healthcare Diagnostics, Inc.

Key facts

Event ID 76695
Recall numbers Z-1712-2017
Recalling firm Siemens Healthcare Diagnostics, Inc.
Product type Devices
Products ADVIA Chemistry Enzyme 2 Calibrator, multi-analyte mixture Product Usage: Assays included in Enzyme 2 Calibrator: AST, AST_c ALT, ALT_c, ALTP5P, ALTP_c, ASTP5P and ASTP_c
Classification Class III
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall Siemens Healthcare Diagnostics has confirmed a positive bias of up to 41% in quality control and patient results for Aspartate Aminotransferase (AST) and Aspartate Aminotransferase concentrated…
Initiation date Feb 23, 2017
Report date Apr 12, 2017
Termination date May 9, 2018
City Newark
State DE
Country United States
FEI number 2517506

At a glance

  • 1–9 Products in this recall
  • None FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date: