Drug recall

Plum 360 Infusion Pump, Software Version 15.02. The infusion pump is capable of delivering fluids for a…

FDA recall event 76253 · Class II · Hospira Inc., A Pfizer Company

Key facts

Event ID 76253
Recall numbers Z-1682-2017
Recalling firm Hospira Inc., A Pfizer Company
Product type Devices
Products Plum 360 Infusion Pump, Software Version 15.02. The infusion pump is capable of delivering fluids for a variety of therapies such as parenteral, enteral, or epidural infusions
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall Depleted Battery alarm shows instead of Replace Battery. On battery power, ongoing therapy stops, alarms show and sound, pump shuts down after 3 minutes. On AC power, the pump reboots (takes about 25…
Initiation date Dec 30, 2016
Report date Apr 5, 2017
Termination date Jul 9, 2019
City Lake Forest
State IL
Country United States
FEI number 3005579246

At a glance

  • 1–9 Products in this recall
  • None FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls