Drug recall

SOMATOM Perspective, Computed tomography x-ray system

FDA recall event 75860 · Class II · Siemens Medical Solutions USA, Inc

Key facts

Event ID 75860
Recall numbers Z-0919-2017, Z-0920-2017, Z-0921-2017
Recalling firm Siemens Medical Solutions USA, Inc
Product type Devices
Products SOMATOM Perspective, Computed tomography x-ray system, SOMATOM Scope, Computed tomography x-ray system, SOMATOM Scope Power, Computed tomography x-ray system
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall Siemens is providing a new installation of the Residual Current Monitor (RCM) in the Line Connection Box (LCB) to resolve a potential defect that could lead to a power cutout during use of the SOMATOM…
Initiation date Nov 22, 2016
Report date Jan 4, 2017
Termination date May 30, 2017
City Malvern
State PA
Country United States
FEI number 2240869

At a glance

  • 1–9 Products in this recall
  • None FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date: