Drug recall

Tosoh Bioscience HLC-723G8 Analyzer for HbA1c testing.

FDA recall event 75727 · Class II · Tosoh Bioscience, Inc.

Key facts

Event ID 75727
Recall numbers Z-1225-2017
Recalling firm Tosoh Bioscience, Inc.
Product type Devices
Products Tosoh Bioscience HLC-723G8 Analyzer for HbA1c testing.
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall The device defect associated with this recall may result in HbA1c values with a negative bias in patients with the hemoglobin variants C, D, E and S. The negative bias occurs at the higher levels of…
Initiation date Nov 7, 2016
Report date Mar 1, 2017
Termination date Feb 9, 2021
City South San Francisco
State CA
Country United States
FEI number 1000123732

At a glance

  • 1–9 Products in this recall
  • 1–9 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date: