Drug recall

FlexLab, Accelerator a3600 and Aptio Automation are modular systems designed to automate Pre-Analytical and…

FDA recall event 75513 · Class II · Inpeco S.A.

Key facts

Event ID 75513
Recall numbers Z-0587-2017
Recalling firm Inpeco S.A.
Product type Devices
Products FlexLab, Accelerator a3600 and Aptio Automation are modular systems designed to automate Pre-Analytical and Post-Analytical processing, sample handling in order to automate sample processing in the…
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall Inpeco received several notifications from the field about burnt schuko sockets. The customers reported that some smoke was seen coming from the damaged sockets.
Initiation date Oct 4, 2016
Report date Nov 23, 2016
Termination date May 31, 2019
City Lugano
Country Switzerland
FEI number 3010825766

At a glance

  • 1–9 Products in this recall
  • None FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date: