Drug recall
Replacement floor locking cylinders for the Schaerer model 7300 surgical tables. Replacement cylinders are…
FDA recall event 75288 · Class II · Schaerer Medical USA Inc
Key facts
| Event ID | 75288 |
|---|---|
| Recall numbers | Z-0371-2017, Z-0372-2017 |
| Recalling firm | Schaerer Medical USA Inc |
| Product type | Devices |
| Products | Replacement floor locking cylinders for the Schaerer model 7300 surgical tables. Replacement cylinders are part number 85118. Product Usage: The cylinders are used to lock the surgical table in place., The surgical table is a stand alone unit. Three (3) floor locking cylinders are attached to the base of the surgical table. The individual cylinders are replaceable. Product Usage: The cylinders are… |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Firm Initiated |
| Reason for recall | The floor locking cylinders are contained in the base of the Schaerer 7300 model surgical tables. The cylinders were manufactured with the incorrect "O" ring which may cause the cylinder to leak., The floor locking cylinders contained in the base of the Schaerer 7300 model surgical tables were manufactured with the incorrect "O" ring which may cause the cylinder to leak. |
| Initiation date | Sep 27, 2016 |
| Report date | Nov 16, 2016 |
| Termination date | Jul 17, 2020 |
| City | Cincinnati |
| State | OH |
| Country | United States |
| FEI number | 1525725 |
At a glance
- 1–9 Products in this recall
- 1–9 FDA inspections of this firm (FEI)
Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.
About FDA drug and product recalls
FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.
Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date: