Drug recall

enGen Laboratory Automation System (Product Code ENGEN) configured with Thermo Scientific Centrifuge Module…

FDA recall event 75167 · Class II · Ortho-Clinical Diagnostics

Key facts

Event ID 75167
Recall numbers Z-0300-2017
Recalling firm Ortho-Clinical Diagnostics
Product type Devices
Products enGen Laboratory Automation System (Product Code ENGEN) configured with Thermo Scientific Centrifuge Module and TCAutomation Software Version 3.6.1 and Below; IVD. --- Thermo Scientific Centrifuge…
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall Ortho confirmed that it is possible for the mis-association of sample identification and assay results due to an error with Thermo Fisher s centrifuge module. It is possible for mis-associated assay…
Initiation date Sep 7, 2016
Report date Nov 2, 2016
Termination date Jun 15, 2018
City Rochester
State NY
Country United States
FEI number 1000136573

At a glance

  • 1–9 Products in this recall
  • 10–99 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls