Drug recall

LipiFlow thermal Pulsation System, consisting of Console and Activator, Model LFD-1100 The LipiFlow Thermal…

FDA recall event 75163 · Class II · Tearscience, Inc

Key facts

Event ID 75163
Recall numbers Z-0093-2017
Recalling firm Tearscience, Inc
Product type Devices
Products LipiFlow thermal Pulsation System, consisting of Console and Activator, Model LFD-1100 The LipiFlow Thermal Pulsation System is intended for use by physicians as a prescription device for the…
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall Complaints related to an error condition occurring during patient treatment where the FipiFlow System automatically stopped treatment after detecting a problem with the Activator.
Initiation date Sep 13, 2016
Report date Oct 26, 2016
Termination date Feb 9, 2017
City Morrisville
State NC
Country United States
FEI number 3008169506

At a glance

  • 1–9 Products in this recall
  • 1–9 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls