Drug recall

NIO-A / NIO-P is an instant IO access device comprises housing with spring load mechanism that injects a…

FDA recall event 75008 · Class II · Waismed Ltd

Key facts

Event ID 75008
Recall numbers Z-2680-2016
Recalling firm Waismed Ltd
Product type Devices
Products NIO-A / NIO-P is an instant IO access device comprises housing with spring load mechanism that injects a trocar needle into the bone marrow of a patient and allows connection of syringe or IV…
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall WaisMed received a report in which during the operation of the NIO-A, the needle was not released from the device as expected. The device was placed on the floor and after several minutes the needle…
Initiation date Aug 18, 2016
Report date Sep 7, 2016
Termination date Dec 5, 2016
City Rosh Haayin
Country Israel
FEI number 3011989505

At a glance

  • 1–9 Products in this recall
  • None FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls