Drug recall

Trinity Biotech Captia Measles IgM, REF/Product Code 2326060,Test Kit for 96 Determinations, IVD

FDA recall event 74932 · Class III · Clark Laboratories, Inc. (dba,Trinity Biotech USA)

Key facts

Event ID 74932
Recall numbers Z-2782-2016
Recalling firm Clark Laboratories, Inc. (dba,Trinity Biotech USA)
Product type Devices
Products Trinity Biotech Captia Measles IgM, REF/Product Code 2326060,Test Kit for 96 Determinations, IVD
Classification Class III
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall The product lot failed to meet QC acceptance criteria during stability testing. The Calibrator OD was dropping causing the positive control to be invalid (running high out of range).
Initiation date Jul 11, 2016
Report date Sep 21, 2016
Termination date May 9, 2017
City Jamestown
State NY
Country United States
FEI number 1318354

At a glance

  • 1–9 Products in this recall
  • 1–9 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date: