Drug recall
PCR Compano Basic Reader using 100-127V as mains voltage. Model numbers: 732027, 732028, and 732040…
FDA recall event 74920 · Class II · Philips Electronics North America Corporation
Key facts
| Event ID | 74920 |
|---|---|
| Recall numbers | Z-2681-2016 |
| Recalling firm | Philips Electronics North America Corporation |
| Product type | Devices |
| Products | PCR Compano Basic Reader using 100-127V as mains voltage. Model numbers: 732027, 732028, and 732040. Radiological image processing system (Computed Radiography). |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Firm Initiated |
| Reason for recall | A defect component of the Compano reader power supply can cause the power supply to burn and some smoke to leave the housing. |
| Initiation date | Aug 12, 2016 |
| Report date | Sep 7, 2016 |
| Termination date | Jul 15, 2020 |
| City | Andover |
| State | MA |
| Country | United States |
| FEI number | 1218950 |
At a glance
- 1–9 Products in this recall
- None FDA inspections of this firm (FEI)
Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.
About FDA drug and product recalls
FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.
Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date: