Drug recall

Anspach G1 High Speed Electric Handpiece Air Cooled (eG1A) Use for cutting and shaping bone including bones of…

FDA recall event 74447 · Class II · The Anspach Effort, Inc.

Key facts

Event ID 74447
Recall numbers Z-2253-2016
Recalling firm The Anspach Effort, Inc.
Product type Devices
Products Anspach G1 High Speed Electric Handpiece Air Cooled (eG1A) Use for cutting and shaping bone including bones of the cranium and spine.
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall The graphics at the attachment interface indicating locked and unlocked position are reversed.
Initiation date Jun 15, 2016
Report date Jul 27, 2016
Termination date Dec 14, 2016
City Palm Beach Gardens
State FL
Country United States
FEI number 1045834

At a glance

  • 1–9 Products in this recall
  • 1–9 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date: