Drug recall

Stryker FlowPort II Adapter: Model number: 00CAT00778 The FlowPort II Adapter is intended to connect the…

FDA recall event 74070 · Class II · Stryker Corporation

Key facts

Event ID 74070
Recall numbers Z-1910-2016
Recalling firm Stryker Corporation
Product type Devices
Products Stryker FlowPort II Adapter: Model number: 00CAT00778 The FlowPort II Adapter is intended to connect the FlowPort II Cannulas to commercially available, 4mm arthroscopes in surgical procedures.
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall Complaints were received for the Stryker FlowPort II Adapter, and investigation found that the scope sat in the adapter 180 degrees in the wrong direction
Initiation date May 4, 2016
Report date Jun 15, 2016
Termination date Jun 3, 2016
City San Jose
State CA
Country United States
FEI number 2936485

At a glance

  • 1–9 Products in this recall
  • 10–99 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date: