Drug recall

SOMATOM Force, Computed Tomography x-ray system intended to generate and process cross-sectional images of…

FDA recall event 73838 · Class II · Siemens Medical Solutions USA, Inc

Key facts

Event ID 73838
Recall numbers Z-1558-2016
Recalling firm Siemens Medical Solutions USA, Inc
Product type Devices
Products SOMATOM Force, Computed Tomography x-ray system intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission.
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall The Neonate Head protocol with the pediatric kernel Hp38 could result in artefacts and possibly lead to a misdiagnosis (i.e. either non-existing blood or liquid is mimicked in the images or actual…
Initiation date Apr 1, 2016
Report date May 4, 2016
Termination date Apr 17, 2017
City Malvern
State PA
Country United States
FEI number 2240869

At a glance

  • 1–9 Products in this recall
  • None FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls