Drug recall

Reveal LINQ insertable Cardiac Monitor, Model LNQ11. Monitors subcutaneous ECG continuously.

FDA recall event 73494 · Class II · Medtronic Inc.

Key facts

Event ID 73494
Recall numbers Z-1249-2016
Recalling firm Medtronic Inc.
Product type Devices
Products Reveal LINQ insertable Cardiac Monitor, Model LNQ11. Monitors subcutaneous ECG continuously.
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall Medtronic has identified an issue with the sensitivity of an algorithm used in the Reveal LINQ ICM that may prematurely trigger the Recommended Replacement Time (RRT) alert in some devices. Battery…
Initiation date Feb 23, 2016
Report date Apr 6, 2016
Termination date Sep 25, 2017
City Saint Paul
State MN
Country United States
FEI number 2182208

At a glance

  • 1–9 Products in this recall
  • 10–99 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date: