Drug recall

LIFEPAK15 monitor/defibrillator with End-Tidal C02 (EtC02) feature installed. Part numbers: V15-2-xxxxxx…

FDA recall event 73473 · Class II · Physio-Control, Inc.

Key facts

Event ID 73473
Recall numbers Z-1993-2016
Recalling firm Physio-Control, Inc.
Product type Devices
Products LIFEPAK15 monitor/defibrillator with End-Tidal C02 (EtC02) feature installed. Part numbers: V15-2-xxxxxx (includes software version 1 and version 2); V15-5-xxxxxx (inclues software version 4). The…
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall The firm became aware that when using EtC02 in the kPa or % setting and in a situation where the reading is above 9.9 kPa, the display of the LIFEPAK 15 respiratory rate may partially obscure a…
Initiation date Jan 27, 2016
Report date Jun 22, 2016
Termination date Mar 31, 2017
City Redmond
State WA
Country United States
FEI number 3015876