1) VITROS¿ Immunodiagnostic Products Total Beta-hCG II Reagent Pack (Unique Device Identifier No…
FDA recall event 73441 · Class II · Ortho-Clinical Diagnostics
Key facts
| Event ID | 73441 |
|---|---|
| Recall numbers | Z-1700-2016 |
| Recalling firm | Ortho-Clinical Diagnostics |
| Product type | Devices |
| Products | 1) VITROS¿ Immunodiagnostic Products Total Beta-hCG II Reagent Pack (Unique Device Identifier No. 10758750002320, 20758750002327) and the 2)VITROS¿ Immunodiagnostic Products Total Beta-hCG II… |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Firm Initiated |
| Reason for recall | When testing was conducted using certain lots, VITROS¿ Systems generated results within the measuring range for samples known to not contain measurable hCG. The system reported results up to… |
| Initiation date | Feb 18, 2016 |
| Report date | Jun 1, 2016 |
| Termination date | May 30, 2018 |
| City | Rochester |
| State | NY |
| Country | United States |
| FEI number | 1000136573 |