VITROS 5600 Integrated System, Unique Device Identifier No. 10758750002740 & 10758750007110. Intended for use…
FDA recall event 73429 · Class II · Ortho-Clinical Diagnostics
Key facts
| Event ID | 73429 |
|---|---|
| Recall numbers | Z-1348-2016 |
| Recalling firm | Ortho-Clinical Diagnostics |
| Product type | Devices |
| Products | VITROS 5600 Integrated System, Unique Device Identifier No. 10758750002740 & 10758750007110. Intended for use in the in vitro measurement of a variety of analytes of clinical interest. |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Firm Initiated |
| Reason for recall | The two sets of wires (4 total wires) exiting from the EMI filters were switched, creating a situation where a module that is intended to be de-energized will remain energized only if the associated… |
| Initiation date | Feb 16, 2016 |
| Report date | Apr 13, 2016 |
| Termination date | May 30, 2018 |
| City | Rochester |
| State | NY |
| Country | United States |
| FEI number | 1000136573 |