Drug recall

VITROS 5600 Integrated System, Unique Device Identifier No. 10758750002740 & 10758750007110. Intended for use…

FDA recall event 73429 · Class II · Ortho-Clinical Diagnostics

Key facts

Event ID 73429
Recall numbers Z-1348-2016
Recalling firm Ortho-Clinical Diagnostics
Product type Devices
Products VITROS 5600 Integrated System, Unique Device Identifier No. 10758750002740 & 10758750007110. Intended for use in the in vitro measurement of a variety of analytes of clinical interest.
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall The two sets of wires (4 total wires) exiting from the EMI filters were switched, creating a situation where a module that is intended to be de-energized will remain energized only if the associated…
Initiation date Feb 16, 2016
Report date Apr 13, 2016
Termination date May 30, 2018
City Rochester
State NY
Country United States
FEI number 1000136573

At a glance

  • 1–9 Products in this recall
  • 10–99 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls