Drug recall

Synaptive Medical BrightMatter Navigator System (Guide System). Part number SYN-0026. The Guide system is…

FDA recall event 72747 · Class II · Synaptive Medical, Inc.

Key facts

Event ID 72747
Recall numbers Z-0480-2016
Recalling firm Synaptive Medical, Inc.
Product type Devices
Products Synaptive Medical BrightMatter Navigator System (Guide System). Part number SYN-0026. The Guide system is comprised of a cart labelled SYN-0024 and a cart labelled SYN-0025, Operator Cart. A planning…
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall Out of tolerance for radio frequency emissions. At the 150-1000MHz frequency, the testing indicated the BrightMatter Navigation system was up to 20dB uV/meter higher than the applicable IEC…
Initiation date Dec 4, 2015
Report date Dec 30, 2015
Termination date Mar 4, 2016
City Toronto
Country Canada
FEI number 3010439744

At a glance

  • 1–9 Products in this recall
  • None FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls