Drug recall

ADVIA Chemistry Hemoglobin A1c_3, A1c_3M Reagent kits. Used on the ADVIA 1200, 1650, 1800, 2400, and XPT…

FDA recall event 72392 · Class II · Siemens Healthcare Diagnostics, Inc.

Key facts

Event ID 72392
Recall numbers Z-0340-2016
Recalling firm Siemens Healthcare Diagnostics, Inc.
Product type Devices
Products ADVIA Chemistry Hemoglobin A1c_3, A1c_3M Reagent kits. Used on the ADVIA 1200, 1650, 1800, 2400, and XPT Chemistry Systems. For in vitro diagnostic use in the quantitative determination of Hemoglobin…
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall Reagent in these lots may demonstrate an increased occurrence of high %HbA1c bias, a positive bias averaging 0.6% HbA1c units, ranging from -0.1% to 1.1% HbA1c units. The maximum bias was observed at…
Initiation date Sep 24, 2015
Report date Dec 2, 2015
Termination date Jan 13, 2017
City Newark
State DE
Country United States
FEI number 2517506

At a glance

  • 1–9 Products in this recall
  • None FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls