Drug recall

ADVIA Chemistry XPT Chemistry Systems, Siemens Material Number (SMN) 10723034. An automated, clinical…

FDA recall event 72104 · Class II · Siemens Healthcare Diagnostics, Inc.

Key facts

Event ID 72104
Recall numbers Z-0183-2016
Recalling firm Siemens Healthcare Diagnostics, Inc.
Product type Devices
Products ADVIA Chemistry XPT Chemistry Systems, Siemens Material Number (SMN) 10723034. An automated, clinical chemistry analyzer that runs tests on serum, plasma, urine, or cerebral spinal fluid in random…
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall ISE module may produce discrepant, unflagged results for Sodium, Potassium or Chloride when the respective electrode is left unplugged after replacement, washing or maintenance activities. The…
Initiation date Aug 28, 2015
Report date Nov 4, 2015
Termination date May 22, 2018
City Tarrytown
State NY
Country United States
FEI number 2432235

At a glance

  • 1–9 Products in this recall
  • None FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls