Drug recall
Neptune 2 Waste Management System Docking Station (120V and 230V), New and Refurbished Product Usage: The…
FDA recall event 71872 · Class II · Stryker Instruments Div. of Stryker Corporation
Key facts
| Event ID | 71872 |
|---|---|
| Recall numbers | Z-2524-2015 |
| Recalling firm | Stryker Instruments Div. of Stryker Corporation |
| Product type | Devices |
| Products | Neptune 2 Waste Management System Docking Station (120V and 230V), New and Refurbished Product Usage: The Stryker Docking Station (Docker) is a component of the Neptune 2 Waste Management System. The… |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Firm Initiated |
| Reason for recall | The Fluid Coupling Sleeve, a component within the Neptune Waste Management System Docker, may be out of specification. This can cut or dislodge the PTFE seal and cause a leak between the Rover and… |
| Initiation date | Jul 20, 2015 |
| Report date | Sep 9, 2015 |
| Termination date | Nov 7, 2016 |
| City | Portage |
| State | MI |
| Country | United States |
| FEI number | 1811755 |
At a glance
- 1–9 Products in this recall
- 10–99 FDA inspections of this firm (FEI)
Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.
About FDA drug and product recalls
FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.
Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date: