Drug recall

Neptune 2 Waste Management System Docking Station (120V and 230V), New and Refurbished Product Usage: The…

FDA recall event 71872 · Class II · Stryker Instruments Div. of Stryker Corporation

Key facts

Event ID 71872
Recall numbers Z-2524-2015
Recalling firm Stryker Instruments Div. of Stryker Corporation
Product type Devices
Products Neptune 2 Waste Management System Docking Station (120V and 230V), New and Refurbished Product Usage: The Stryker Docking Station (Docker) is a component of the Neptune 2 Waste Management System. The…
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall The Fluid Coupling Sleeve, a component within the Neptune Waste Management System Docker, may be out of specification. This can cut or dislodge the PTFE seal and cause a leak between the Rover and…
Initiation date Jul 20, 2015
Report date Sep 9, 2015
Termination date Nov 7, 2016
City Portage
State MI
Country United States
FEI number 1811755

At a glance

  • 1–9 Products in this recall
  • 10–99 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls