Drug recall

Philips DigitalDiagnost Release 4.0.3, Release 4.1/4.1.1. excluding systems with SP1 Product Usage: The…

FDA recall event 71838 · Class II · Philips Electronics North America Corporation

Key facts

Event ID 71838
Recall numbers Z-2383-2015
Recalling firm Philips Electronics North America Corporation
Product type Devices
Products Philips DigitalDiagnost Release 4.0.3, Release 4.1/4.1.1. excluding systems with SP1 Product Usage: The DigitalDiagnost is intended for use in generating radiographic images of human anatomy by…
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall The system is designed to emit a beep upon termination of an exposure. However, if the system has been powered on for over 12 hours, the system will no longer emit this signal due to a defect in the…
Initiation date Mar 23, 2015
Report date Sep 2, 2015
Termination date Jan 17, 2017
City Andover
State MA
Country United States
FEI number 1218950

At a glance

  • 1–9 Products in this recall
  • None FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls