Drug recall

OPD-Scan III Refractive Power/Corneal Analyzer Opthalmic Software versions 1.00.08 1.05.07 1.10.01, 1.01.02…

FDA recall event 71691 · Class II · Nidek Inc

Key facts

Event ID 71691
Recall numbers Z-2477-2015
Recalling firm Nidek Inc
Product type Devices
Products OPD-Scan III Refractive Power/Corneal Analyzer Opthalmic Software versions 1.00.08 1.05.07 1.10.01, 1.01.02 1.06.02 1.11.02, 1.02.01 1.07.01 1.12.03, 1.03.02 1.08.01 1.13.01, 1.04.03 1.09.01…
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall Software bug was found where there was no difference in Total and Corneal high-order aberrations, but differences were found in Internal high-order aberrations.
Initiation date Jun 15, 2015
Report date Sep 2, 2015
Termination date Apr 25, 2016
City Fremont
State CA
Country United States
FEI number 2936921

At a glance

  • 1–9 Products in this recall
  • 10–99 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date: