Drug recall

GE Healthcare Precision 500D and Advantx Legacy/Legacy-D Radiographic and Fluoroscopic Systems. A universal…

FDA recall event 71572 · Class II · GE Healthcare

Key facts

Event ID 71572
Recall numbers Z-1966-2015
Recalling firm GE Healthcare
Product type Devices
Products GE Healthcare Precision 500D and Advantx Legacy/Legacy-D Radiographic and Fluoroscopic Systems. A universal diagnostic imaging system for radiographic and fluoroscopic examinations. Applications can…
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall The SFD/IDD safety mechanism may not engage properly at the lower range of SFD/IDD travel. This mechanism is installed to hold and prevent the SFD/IDD from falling due to a counterweight cable…
Initiation date Jun 12, 2015
Report date Jul 8, 2015
Termination date Dec 21, 2015
City Waukesha
State WI
Country United States
FEI number 2126677

At a glance

  • 1–9 Products in this recall
  • 10–99 FDA inspections of this firm (FEI)