Radiation Therapy Treatment Planning System software RayStation is a software system designed for treatment…

FDA recall event 71477 · Class II · RAYSEARCH LABORATORIES AB

Key facts

Event ID 71477
Recall numbers Z-2785-2015
Recalling firm RAYSEARCH LABORATORIES AB
Product type Devices
Products Radiation Therapy Treatment Planning System software RayStation is a software system designed for treatment planning and analysis of radiation therapy. The treatment plans provide treatment unit…
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall An error caused by a floating point precision problem tends to cause a problem with the display of isodose lines for dose and dose difference and color tables for dose, dose difference and PET images
Initiation date May 29, 2015
Report date Sep 23, 2015
Termination date Mar 13, 2017
City Stockholm
Country Sweden
FEI number 3007774465