Magnus Hybrid OR Table Column (1180.01B2) used in conjunction with Siemens Artis Zeego Angiography System…
FDA recall event 71446 · Class II · Maquet Cardiovascular Us Sales, Llc
Key facts
| Event ID | 71446 |
|---|---|
| Recall numbers | Z-2097-2015 |
| Recalling firm | Maquet Cardiovascular Us Sales, Llc |
| Product type | Devices |
| Products | Magnus Hybrid OR Table Column (1180.01B2) used in conjunction with Siemens Artis Zeego Angiography System. Maquet GmbH & Co. |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Firm Initiated |
| Reason for recall | There is a potential issue that could result to collision between the C-arm of the angiography system and the MAGNUS OR table system. After switching on the angiography-system, in rare cases, MAGNUS… |
| Initiation date | Jun 10, 2015 |
| Report date | Jul 29, 2015 |
| Termination date | Apr 18, 2016 |
| City | Wayne |
| State | NJ |
| Country | United States |
| FEI number | 3008355164 |