Drug recall

CareFusion AVEA ventilator, all models. Product Usage: A continuous ventilator (respirator), intended to…

FDA recall event 70920 · Class I · Carefusion 211 Inc dba Carefusion

Key facts

Event ID 70920
Recall numbers Z-1609-2015
Recalling firm Carefusion 211 Inc dba Carefusion
Product type Devices
Products CareFusion AVEA ventilator, all models. Product Usage: A continuous ventilator (respirator), intended to mechanically control or assist patient breathing by delivering a predetermined percentage of…
Classification Class I
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall A pressure transducer failure can develop, activating a false Extended High Ppeak or Circuit Occlusion Alarm. The Safety Valve will open to ambient air and the unit will stop ventilating, allowing…
Initiation date Apr 3, 2015
Report date Jun 3, 2015
Termination date Jul 23, 2018
City Yorba Linda
State CA
Country United States
FEI number 2050001

At a glance

  • 1–9 Products in this recall
  • None FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls